Pressure Systems, Inc. NetScanner™ System (9016, 9021, & 9022) User’s Manual
www.PressureSystems.com
3 - 29
Example:
! Send TCP/IP command to 9016 module (via its connected socket) that returns
decimal raw “pressure” A/D counts data for channels 1, 5, 9, and 13:
a11110
Response contains data for channels 13, 9, 5, and 1 (left to right):
32767.000000 -32700.000000 10.000000 16385.000000
Please note that channel 13 is saturated at +full scale and channel 9 is almost
saturated at -full scale. Channel 5 reads near zero and channel 1 is about ½ +full-
scale.
This command example also works for Models 9021 and 9022 if the non-existent
channel 13 is not set in the position field bit map.