Measurement Specialties Inc. NetScanner System (9016, 9021, & 9022) User’s Manual
3 - 16
CONFIGURE/CONTROL MULTI-POINT CALIBRATION (Command ‘C’)
Purpose: This command is actually four (4) sub-commands. The first configures and starts
a Multi-Point Calibration adjustment function for selected channels in the
module. Another is repeated multiple times to collect data for each defined
calibration point. Another ends the calibration function normally by calculating
new offset and gain adjustment coefficients from the collected data. It then
returns the module to its normal state, but with improved accuracy. A final
sub-command is used only if it becomes necessary to abort the calibration
function while in progress. The general form of all sub-commands is described
in the table below. Subsequent pages separately describe each individual
sub-command and give examples of each.
Command
C ii[ dddd]...
‘Cis the command letter.
ii’ is a required sub-command index preceded by a space character.
dddd’ are zero or more optional datum (or parameter) fields, each preceded by
a space character. These vary with the sub-command used.
Response
Depends upon the particular sub-command (ii) used.
Description: The four C’ sub-commands configure and control operation of a Multi-Point
Calibration function that is similar to the simpler re-zero and span calibration
adjustment functions (see separate ‘h’ and Z’ commands). However, ‘C’ adjusts
both the offset and gain correction coefficients of each affected transducer at the
same time, using two or more calibration points. Thus, instead of simply
calculating a new linear (i.e., straight line) adjustment function that passes
through the supplied zero and span calibration points, it calculates a best-fit
straight line, using the least squares method, that comes “as close as possible” to
all the supplied calibration points. This correction method provides the very
best adjustment throughout the entire range (negative to positive) of a differential
transducer.
This ‘C’ command (with sub-commands) is available only in
modules that have upgraded to firmware Version 2.24 or later.