DOWNLOAD INTERNAL COEFFICIENTS (Command ‘v’)
Purpose: Downloads one or more internal coefficients to the module.
Command
“vfaacc[-cc]dddd..dddd”
‘v’ is the command letter.
‘f’ is the format field.
‘aa’ is the array index field.
‘cc[-cc]’ is coefficient index [or contiguous range].
‘dddd’ are the datum field(s) each with a leading
space.
Response
“A”
‘A’ is the acknowledge letter.
Description: The 1-character format field (f) is a single decimal digit that defines the format of
each coefficient to be downloaded in the command’s datum ( dddd) fields, with
each datum preceded by a space character. Most coefficients have a floating
point datum type (f=0-1), while others have an integer datum type (f=5). Sending a
datum in the improper format will result in an “N08" error response. Valid format
types are shown in the following table:
f
converts each datum parameter value (‘ dddd’) from..
max. char.
0
1-10 digit signed decimal “ [-xxx]x.[xxxxxx]” to single
binary float
13
1
8-digit hex “ xxxxxxxx”
to
single
binary float
9
5
8-digit hex “ xxxxxxxx”
to
long
binary integer
9
The 2-character array index field (aa) is a hexadecimal value selecting a
particular internal coefficient array to receive the downloaded data. The first
array index (aa=01) refers to channel 1's transducer, the 16th (aa=10) refers to
channel 16's transducer. Finally, the last array (aa=11) refers to a special global
array.