Pressure Systems 9022 Uživatelský manuál Strana 62

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 194
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 61
www.meas-
spec.com
Measurement Specialties Inc. NetScanner System (9016, 9021, & 9022) User’s Manual
3 - 29
Example:
Send TCP/IP command to 9016 module (via its connected socket) that returns
decimal raw “pressure” A/D counts data for channels 1, 5, 9, and 13:
a11110
Response contains data for channels 13, 9, 5, and 1 (left to right):
32767.000000 -32700.000000 10.000000 16385.000000
Please note that channel 13 is saturated at +full scale and channel 9 is almost
saturated at -full scale. Channel 5 reads near zero and channel 1 is about ½
+full-scale.
This command example also works for Models 9021 and 9022 if the non-existent
channel 13 is not set in the position field bit map.
Zobrazit stránku 61
1 2 ... 57 58 59 60 61 62 63 64 65 66 67 ... 193 194

Komentáře k této Příručce

Žádné komentáře